Remote control commands
R&S
®
SMW-K553
37
User Manual 1179.4240.02 ─ 02
Example:
See
Example "Calibrating external frontends"
Usage:
Event
[:SOURce<hw>]:EFRontend:CONNection:STATe
<ConnState>
Queries the state of the connection between R&S
SMW and external frontend.
Parameters:
<ConnState>
1 | ON | 0 | OFF
*RST:
0
Example:
See
Example "Configuring external frontend connection set-
[:SOURce<hw>]:EFRontend:CMODe
<FeConnMode>
Sets the mode of the SSL control connection between R&S
SMW and external fron-
tend.
Parameters:
<FeConnMode>
AUTO | LOCK | RXTX
AUTO
The R&S
SMW locks external frontend, when activating the RF
output at the R&S
SMW ("RF" > "On") for output of the IF signal.
The R&S
SMW unlocks external frontend, when deactivating the
RF output at the R&S
SMW ("RF" > "Off").
LOCK
The external frontend is locked permanently. No other instru-
ment can take over control.
RXTX
Requires an R&S
FE50DTR connected to the R&S
SMW.
Connection mode for simultaneous Rx (receive) operation and
Tx (transmit) operation of an R&S
FE50DTR.
Set this mode, if your test setup requires a release of connection
lock. A connected vector signal analyzer can lock the external
frontend for Rx operation.
Example:
See
Example "Configuring external frontend connection set-
Manual operation:
See
[:SOURce<hw>]:EFRontend:DID?
Queries the device ID of the external frontends connected to the R&S
SMW.
Return values:
<FEDevID>
string
<product_name>-<SerialNumber>
SOURce:EFRontend subsystem