Remote control commands
R&S
®
SMW-K553
42
User Manual 1179.4240.02 ─ 02
NONE
No frontend connected.
RFB
Fixed output connector "RF B" at the external frontend.
Example:
See
Example "Configuring external frontend connection set-
Manual operation:
See
6.2
CALibration subsystem
The
CALibration
subsystem contains the commands for calibration of an RF fron-
tend.
Commands:
:CALibration<hw>:FREQuency:CONVerter:EXTernal?
Queries the calibration state of the connected external instrument.
External instruments can be an I/Q upconverter R&S
SZU or an external frontend, e.g.,
R&S
FE44S.
Return values:
<Success>
1 | ON | 0 | OFF
*RST:
0
Example:
// Query calibration state of a connected R&S SZU.
CAL1:FREQ:CONV:EXT?
// 1
// The R&S SZU connected in path A is calibrated.
// Query calibration state of a connected external frontend.
CAL1:FREQ:CONV:EXT?
// 1
// The external frontend connected in path A is calibrated.
Usage:
Query only
CALibration subsystem