R&S SM300
Module Registration Software
Version 04
63
Service Manual SM300: 1147.1652.00
Instrument Part No SM300: 1147.1498.03
Series 300 wizard
for modifying
instrument data
The Series 300 Wizard is a utility for modifying data stored on the instrument.
It is activated by clicking the “Write to the Instrument Wizard” button (see
“Control buttons” above). The user is guided step by step through entering
the data. The number of steps and required information depend on the action
selected (see “Wizard start window” below). The last step is to store data on
the instrument. The user may repeat this procedure as many times as
necessary.
Wizard start window
Description of the actions:
Backplane exchange
Choose this if the backplane was replaced. You will be asked to enter the
serial number of the new backplane, so keep in mind to obtain it and
write it down before starting the application.
Keyboard exchange
Choose this if the front unit containing the keyboard was replaced. You
will be asked to enter the serial number of the new keyboard, so keep in
mind to obtain it from the label on the box and write it down before
starting the application.
Measurement module exchange
Choose this if the measurement module was replaced.
PC module exchange
Choose this if the PC module was replaced.
Power supply exchange
Choose this if the power supply was replaced. When replacing the power
supply, the data from the old unit has to be read out, stored to a file. To
do this, you need to connect the instrument (still containing the old power
supply) to the USB, run the application, read out data from the instrument
and store it to a file (see section “User Interface Description”). If the
power supply is so badly damaged that this cannot be done, it is
important to read the instrument, backplane and keyboard serial
numbers from the instrument and write them down before starting the
application.
Other action
Choose this if none of above listed actions describes the modification
you made on the instrument. This gives you the possibility to enter the
modification description for the service circular, as well as the service
information, and store them on the instrument.