Instrument Assembly for Board and Module Exchange
R&S SM300
Version 04
46
Service Manual SM300: 1147.1652.00
Instrument Part No SM300: 1147.1498.03
1.3.3.11
Generator Module SM300
Removing generator
module SM300
1.
Remove both protectors (see section 1.3.3.1).
2.
Remove handle (see section 1.3.3.2).
3.
Remove rear frame and both cases (see section 1.3.3.3).
4.
Remove carrier lid (see section 1.3.3.5).
5.
Remove duoblind (see section 1.3.3.8).
6.
Pull generator module out toward rear of instrument. When removing
the module, make sure that the springs in the carrier unit are not
damaged (see Figure 1-30).
Figure 1-30
Installing generator
module SM300
1.
Insert generator module from the top, push it into the instrument until it
connects to backplane. When installing the module, make sure that the
springs in the carrier unit are not damaged.
2.
Install duoblind (see section 1.3.3.8).
3.
Install carrier lid (see section 1.3.3.5).
4.
Install both cases and rear frame (see section 1.3.3.3).
5.
Install handle (see section 1.3.3.2).
6.
Install both protectors (see section 1.3.3.1).