Replacing Modules
R&S SM300
Version 04
56
Service Manual SM300: 1147.1652.00
Instrument Part No SM300: 1147.1498.03
Troubleshooting
If the SM300 is still in remote mode after restart it could be due to one of the
following:
The USB cable has not been removed from the instrument.
There could be a problem with the USB connection (backplane,
module) that is preventing the control PC from communicating with the
generator module.
A wrong software version has been installed, please check version of
the *.out file.
An error occurred during download and the software was not correctly
installed. In this case, please repeat steps 5 to 12.
Registering module
and verifying
instrument
If you have not run the “Module Registration Software” as described in
section 1.5, please do it now. Otherwise the replaced generator module is
not registered correctly within the instrument.
Once the instrument has been completely updated and reassembled, please
follow the instructions in section 1.6 to ensure correct functionality of the
whole instrument.