R&S OSP
Instrument Functions
Operating Manual 1505.3896.12 - 14
103
Figure 5-41: Block diagram Module R&S OSP-B129
When the module OSP-B129 is configured within the OSP120, a manual operation of
the module can be performed via the OSP Panel. See
chapter 3.3
For remote control operation see
chapters 6
and
The OSP-B129 is a one-slot module. Within one OSP120 up to three modules of this
type can be installed.
The OSP frame offers a maximum of three slots. If more than one OSP-B129 are
required, the extension unit OSP150 must be used. For further information on the
extension unit OSP150 see
chapter 8.2
The description above applies to further options of the R&S OSP as far as the option
has got a similar relay configuration. It will apply to the following modules:
Option
Relay configuration
OSP-B119
1 x SP8T relay, non terminated
2 x SPDT relay, non terminated
5.2.14 Power Sensor Module R&S OSP-PM-I
The module R&S OSP-PM-I allows to integrate a Power Sensor within the R&S OSP.
The module R&S OSP-PM-I includes one N-type feed-through connector and one USB
Adapter, both mounted in the modules front panel.
The module needs to be completed with a Power Sensor of the R&S family NRP-Zxx.
It is recommended to use a power sensor version with short cable (for example R&S
NRP-Z211 with Id. No. 1417.0409.04).
The module R&S OSP-PM-I is designed with base plate, where the Power Sensor
together with cable is fitted. The Power Sensor with its N-type connector is directly
screwed to the modules N-type feed-through connector. The Power Sensor is
controlled via USB interface. The Power Sensors USB connector (Odumac L-series
with 6 pins) is connected to the USB Adapter. The USN adapter serves as filter at the
same time.