![R&S OSP Скачать руководство пользователя страница 125](http://html1.mh-extra.com/html/rands/osp/osp_operating-manual_3076499125.webp)
R&S OSP
Instrument Functions
Operating Manual 1505.3896.12 - 14
110
running some browser (Internet Explorer or Mozilla Firefox for example) on this
computer. Type the following as URL:
http://192.168.48.147/vnc (example with R&S OSP default IP address, see
chapter 2.9
for more information on IP address).
In the following dialog you are prompted for a password. Type instruments as
password.
When finished the start screen displays some information on the equipment, as
described in the following section.
The following keystrokes on the external keyboard correspond to the keys on the front
panel of the R&S OSP130:
key on R&S OSP130 front panel
corresponding keyboard key
MENU
Top level menu
BACK
up in hierarchy by one menu level
HOME
Start screen
CTRL + M
ESC
CTRL + H
cursor keys
cursor keys or mouse
FUNCTION
switching action, if applicable.
OK confirms a selection
STATUS
status information or changes selection of
a switch for path configuration.
CTRL + F
Enter, or mouse click
CTRL + S
5.3.1.1 Starting the R&S OSP120 GUI without password
For some applications it can be reasonable to start the R&S OSP120 GUI without
password, for example if you use the R&S OSP120 within a system that does already