Remote Control
Operating Manual 1505.3896.12 - 14
135
: Opening new VISA channel: TCPIP0::10.123.10.173::inst0::INSTR
: [-->TCPIP0::10.123.10.173::inst0::INSTR] Setting timeout to 5000 ms
: Connection to TCPIP0::10.123.10.173::inst0::INSTR established!
: Session handle: 0
: VISA Resource-Identifier: TCPIP0::10.123.10.173::inst0::INSTR
: send_Query(0, "*IDN?")
: [-->TCPIP0::10.123.10.173::inst0::INSTR] *IDN?
: read_Answer(0, ..., False)
: [<--TCPIP0::10.123.10.173::inst0::INSTR]
Rohde&Schwarz,OSP130,100001,1.0
: send_Query(0, "system:error?")
: [-->TCPIP0::10.123.10.173::inst0::INSTR] system:error?
: read_Answer(0, ..., False)
: [<--TCPIP0::10.123.10.173::inst0::INSTR] 0,"No error"
6.1.2 Switchover to Remote Control
On power-up, the instrument is always in the manual operating state and can be
operated via the front panel controls (for instruments equipped with a display
–
R&S OSP130), a connected keyboard or via the Graphical User Interface (GUI)
displayed on an external monitor. The instrument is switched to remote control as soon
as it receives a command from the controller.
6.1.3 Return to Manual Operation
Return to manual operation can be initiated via the front panel or via remote control.
●
Manually: Click a front panel key, and select the desired action.
●
Via VXI-11 protocol: >L interface message
6.2 Messages
The messages transferred between the controller and the R&S OSP can be either
interface messages or Device Messages (Commands and Device Responses).
Chapter 6.2.3
, SCPI Command Structure and Syntax, describes the structure of the
device messages as defined by the SCPI standard.
6.2.1 VXI-11 Interface Messages
The VXI-11 protocol allows the instrument to be controlled in a Local Area Network.
For a short introduction and a list of interface functions refer to the VXI-11 Protocol
Specification document.