Remote Control Commands
R&S
®
NRPxxS(N)
115
User Manual 1177.5079.02 ─ 10
Parameters:
<magnitude>
0.0
≙
ideal matched source
1.0
≙
total reflection
Range:
0.0 to 1.0
*RST:
0.0
Manual operation:
See
[SENSe<Sensor>:]SGAMma:PHASe
<phase>
Sets the phase angle of the complex reflection coefficient of the source,
Γ
source
.
Parameters:
<phase>
Range:
-360.0 to 360.0
*RST:
0.0
Manual operation:
See
9.8.4.5
Using the S-Parameters Tool
The S-Parameters tool helps loading an S-parameter table into the power sensor. The
S-Parameters tool is part of the R&S NRP Toolkit, see
Chapter 5.1, "R&S NRP Toolkit"
To start the S-Parameters tool
► In the Windows start menu, select "NRP Toolkit" > "S-Parameter Update Multi".
Configuring Basic Measurement Parameters