UltraRAE 3000 User’s Guide
22.6.4 Datalog Type
The instrument has three datalog types:
Auto
Default mode. Collects datalog information when the
instrument is sampling.
Manual
Datalogging occurs only when the instrument’s
datalogging is manually started (see below for
details).
Snapshot
Datalogs only during single-event capture sampling.
Note:
You can only choose one datalog type to be active at a time.
1.
Press [N/-] to step from one option to the next.
2.
Press [Y/+] to make your selection (the dark circle in the “radio
button” indicates “on”).
3.
When you have completed your selection, press [MODE].
You will see two choices: Save and Undo. You have the opportunity
to register the new settings or to change your mind and revert to your
previous settings.
•
Press [Y/+] to save the changes.
Press [N/-] to undo the changes and revert to the previous settings.
22.6.5 Manual Datalog
When the instrument is set to Manual Datalog, you turn datalogging
on and off by stepping through the displays from the Main Display,
and then pressing the keys to select datalog on/off functions.
•
When you reach the screen that says “Start Datalog?” press
[Y/+] to start it. You see “Datalog Started,” confirming that
datalogging is now on.
When you reach the screen that says “Stop Datalog?” press [Y/+] to
stop it. You see “Datalog Stopped,” confirming that datalogging is
now off.
76
Содержание UltraRAE 3000
Страница 30: ...UltraRAE 3000 User s Guide Compound Specific Tube Note Dashed line indicates automatic progression 28...
Страница 31: ...UltraRAE 3000 User s Guide 29...
Страница 41: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 39...
Страница 59: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 57...
Страница 79: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 77...
Страница 95: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 93...
Страница 97: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 95...
Страница 120: ...UltraRAE 3000 User s Guide countdown the shutoff operation is canceled and the instrument continues normal operation 118...
Страница 128: ......