UltraRAE 3000 User’s Guide
19.1 Entering Calibration
1.
Press and hold [MODE] and [N/-] until you see the Password
screen.
2.
In Basic User Level, you do not need a password to perform
calibrations. Instead of inputting a password, enter calibration
by pressing [MODE].
Note:
If you inadvertently press [Y/+] and change any of the
numbers, simply press [MODE] and you will be directed to
the calibration menu.
The Calibration screen is now visible with Zero Calibration
highlighted.
These are your options:
•
Press [Y/+] to select the highlighted calibration (Zero Calib
or Span Calib).
•
Press [MODE] to exit calibration and return to the main
display and resume measurement.
•
Press [N/-] to toggle the highlighted calibration type.