UltraRAE 3000 User’s Guide
5.
Press [Y/+] to initiate calibration.
Note:
You may press [MODE] if you decide that you do not
want to initiate calibration. This will take you directly to the
Calibration menu for Zero calibration.
6.
Span calibration starts a 30-second countdown and displays
this message:
Calibrating...
During the Span calibration process, the instrument performs the
Span calibration automatically and does not require any action on
your part.
Note:
If you want to abort the Span calibration process, press [N/-] at
any time during the process. You will see a confirmation message that
says “Span is aborted!” and then the Zero calibration menu will
appear. You can then proceed to perform a Zero calibration, perform
a Span calibration, or exit to the topmost Calibration menu.
When Span calibration is complete, you will see a message similar to
this (the value shown here is for example only):
Span 2 is done!
Reading = 1000 ppm
The instrument then exits Span calibration and shows the Zero
calibration menu on its display.
Note:
The reading should be very close to the span gas value.
20.2 Exiting Three-Point Calibration
When you are done performing calibrations, press [MODE], which
corresponds with “Back” on the display. You will see the following
message:
Updating settings…
59
Содержание UltraRAE 3000
Страница 30: ...UltraRAE 3000 User s Guide Compound Specific Tube Note Dashed line indicates automatic progression 28...
Страница 31: ...UltraRAE 3000 User s Guide 29...
Страница 41: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 39...
Страница 59: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 57...
Страница 79: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 77...
Страница 95: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 93...
Страница 97: ...UltraRAE 3000 User s Guide Note Dashed line indicates automatic progression 95...
Страница 120: ...UltraRAE 3000 User s Guide countdown the shutoff operation is canceled and the instrument continues normal operation 118...
Страница 128: ......