QRAE II
User’s Guide
Page 83
Note:
If you see the following message, check that the gas is flowing and the hose is
attached:
No Gas Flow... Apply gas or hit any key to start.
When calibration is complete, you will see the screen change between these two:
Multiple Sensor Calibrated!
and
CO = Pass
H2S = Pass
O2 = Pass
-------
LEL = Pass
A sensor calibration failure is indicated like this:
CO = Fail
H2S = Pass
O2 = Pass
-------
LEL = Pass
If a sensor fails, try calibrating again. If calibration fails again, replace the sensor.
Single Sensor Calibration
This procedure determines the second point of the sensor calibration curve for a single
sensor. The display counts down from 60 to 0 seconds, and then shows the name of each
sensor, whether it passed or failed calibration, and the calibrated value for each sensor.
Select the sensor to calibrate by pressing [MODE] until the sensor’s name is highlighted.
Then press [Y/+] to toggle the selection on (indicated by an asterisk, *) or off (no
asterisk).
After you have selected the sensor you want calibrated, it is time to perform the
calibration.
1.
Attach the Calibration Adapter to the QRAE II.
2.
Attach the regulator to the Calibration Gas cylinder.
Содержание QRAE II
Страница 2: ...Copyright 2011 RAE Systems Inc...
Страница 35: ...QRAE II User Guide Page 33 The following diagram shows the navigation of this set of submenus...
Страница 39: ...QRAE II User Guide Page 37...
Страница 42: ...QRAE II User Guide Page 40 The diagram below shows the difference between the two startup modes...
Страница 47: ...QRAE II User Guide Page 45 5 Alarm Signal Summary Note Backlight automatically turns on when QRAE II alarms...
Страница 63: ...QRAE II User Guide Page 61 4 Push the screws from the back of the monitor to pop the cover off the front...
Страница 72: ...QRAE II User s Guide Page 70 poisons and contaminants Range Resolution Response Time LEL 0 100 1 15 sec...
Страница 91: ...QRAE II User s Guide Page 89 Alarm Signal Summary Note Backlight automatically turns on when QRAE II alarms...
Страница 92: ...P N 020 4100 000 Rev F May 2013...