QRAE II
User’s Guide
Page 78
2.
Turn on the regulator to start the Calibration Gas flow.
3.
Attach a hose to the Monitor gas inlet and to the regulator.
4.
Press [MODE] until OK is highlighted.
5.
Press [Y/+] to start calibration.
You should see a countdown from 60 seconds:
Calibration
in progress ... 60
Note:
If you see the following message, check that the gas is flowing and the hose is
attached:
No Gas Flow... Apply gas or hit any key to start.
When calibration is complete, you will see the screen change between these two:
Multiple Sensor Calibrated!
and
CO = Pass
H2S = Pass
O2 = Pass
LEL = Pass
A sensor calibration failure is indicated like this:
CO = Fail
H2S = Pass
O2 = Pass
LEL = Pass
If a sensor fails, try calibrating again. If calibration fails again, replace the sensor.
Single Sensor Calibration
This procedure determines the second point of the sensor calibration curve for a single
sensor. The display counts down from 60 to 0 seconds, and then shows the name of each
sensor, whether it passed or failed calibration, and the calibrated value for each sensor.
Select the sensor to calibrate by pressing [MODE] until the sensor’s name is highlighted.
Then press [Y/+] to toggle the selection on (indicated by an asterisk, *) or off (no
asterisk).
After you have selected the sensor you want calibrated, it is time to perform the
calibration.
Содержание QRAE II
Страница 2: ...Copyright 2011 RAE Systems Inc...
Страница 35: ...QRAE II User Guide Page 33 The following diagram shows the navigation of this set of submenus...
Страница 39: ...QRAE II User Guide Page 37...
Страница 42: ...QRAE II User Guide Page 40 The diagram below shows the difference between the two startup modes...
Страница 47: ...QRAE II User Guide Page 45 5 Alarm Signal Summary Note Backlight automatically turns on when QRAE II alarms...
Страница 63: ...QRAE II User Guide Page 61 4 Push the screws from the back of the monitor to pop the cover off the front...
Страница 72: ...QRAE II User s Guide Page 70 poisons and contaminants Range Resolution Response Time LEL 0 100 1 15 sec...
Страница 91: ...QRAE II User s Guide Page 89 Alarm Signal Summary Note Backlight automatically turns on when QRAE II alarms...
Страница 92: ...P N 020 4100 000 Rev F May 2013...