QRAE II User Guide
Page 22
3.
Block the inlet for 3 seconds. “I” (idle) maximum should increase at least 10
counts.
Note:
If the increase is less than 10 counts, the pump is weak.
4.
Write down the blocked reading for “I.”
5.
Add the blocked number and the unblocked number and divide the total by 2 to
arrive at the proper stall value. Here is an equation for this:
Stall = (Idle
max
+ Block
max
) / 2
6.
Set the stall value to this number.
4.4.2
Adjust Pump Stall: Low
This procedure sets the stall value for the pump’s low-speed setting. See page 39 for
directions on changing the pump speed from high to low, and vice versa.
Note:
At low
speed, if the inlet is blocked, the pump should stop working.
1.
In Programming Mode, set the pump to operate at Low speed.
Exit Programming Mode, shut down the QRAE II, and then restart it in
Diagnostic Mode.
In Diagnostic Mode, step through the menus until you come to Pump Low Speed Stall.
•
The display shows idle maximum and minimum values (for example, 40/37),
plus the stall value with the pump operating at its low speed.
Содержание QRAE II
Страница 2: ...Copyright 2011 RAE Systems Inc...
Страница 35: ...QRAE II User Guide Page 33 The following diagram shows the navigation of this set of submenus...
Страница 39: ...QRAE II User Guide Page 37...
Страница 42: ...QRAE II User Guide Page 40 The diagram below shows the difference between the two startup modes...
Страница 47: ...QRAE II User Guide Page 45 5 Alarm Signal Summary Note Backlight automatically turns on when QRAE II alarms...
Страница 63: ...QRAE II User Guide Page 61 4 Push the screws from the back of the monitor to pop the cover off the front...
Страница 72: ...QRAE II User s Guide Page 70 poisons and contaminants Range Resolution Response Time LEL 0 100 1 15 sec...
Страница 91: ...QRAE II User s Guide Page 89 Alarm Signal Summary Note Backlight automatically turns on when QRAE II alarms...
Страница 92: ...P N 020 4100 000 Rev F May 2013...