111
15.3 Alarm Signal Summary
15.3.1 Hygiene Mode
Alarm Type
Buzzer & LED
Display
Vibration
Reading
Backlight
Priority
Super Alarm
4 beeps/sec
“Super Alarm” screen
400ms
-
On
Highest
Man Down
Alarm
3 beeps/sec
“Man Down Alarm”
screen
400ms
-
On
Man Down
Warning
2 beeps/sec
“Are you OK” screen
400ms
-
On
Fail
3 beeps/sec
“Lamp” at PID location
“Off” at LEL location
400ms
Blinking
reading
On
Pump
3 beeps/sec
Blinking pump symbol
400ms
Reading
On
Max
3 beeps/sec
“Max” at sensor location
400ms
Blinking
reading
On
Over Range
3 beeps/sec
“Over” at sensor location
400ms
Blinking
9999
On
High
3 beeps/sec
“High” at sensor location
400ms
Reading
On
Low
2 beeps/sec
“Low” at sensor location
400ms
Reading
On
Negative
1 beep/sec
“Neg” at sensor location
400ms
0
On
STEL
1 beep/sec
“STEL” at sensor
location
400ms
Reading
On
TWA
1 beep/sec
“TWA” at sensor
location
400ms
Reading
On
Calibration
Fail
1 beep/sec
“Cal” at sensor location
400ms
Reading
On
Bump Fail
1 beep/sec
“Bump” at sensor location
400ms
Reading
On
Datalog Full
1 beep/sec
Blinking datalog symbol
400ms
Reading
On
Calibration
Required
-
“Full” Bottle symbol
-
Reading
-
Bump
Required
-
“Empty” Bottle symbol
-
Reading
-
Battery
1 beep/min
Blinking battery symbol
400ms
Reading
Stays as
is
Nwk Lost
1 beep/min
Blinking RF offline
symbol
400ms
Reading
On
Nwk Joined
1 beep
RF symbol with RSSI
400ms
Reading
On
Comfort
Beep
1 beep/min
no LED flash
-
-
Reading
-
Lowest
Notes
“Negative” means that the true sensor reading is below zero, even though a zero reading is
shown for the sensor.
“Nwk Lost” means “Network Lost.” This indicates that the MuliRAE has lost wireless
connectivity with its network.
“Nwk Joined” means that the MultiRAE has joined a wireless network.