117
Hold your finger over the inlet, and allow the countdown to proceed. Press [N/-] anytime to
abort.
When the countdown is finished, the main Dynamic pump stall screen is shown.
Dynamic pump stall calibration is complete. You can now exit Diagnostic Mode.
Important!
When you are done setting thresholds, exit Diagnostic Mode and test the instrument
before actual use.
17.4 Setting Pump Stall Threshold Values
– Static Method
Use the following values for reference when using the Static method to adjust the pump stall
threshold values:
Low Speed
High Speed
Vacuum
≤ -2.5in Hg
≤ -10in Hg
Flow rate
>200 cc/min
>300 cc/min
Idle (I)
100 to 200 counts
150 to 250 counts
Block
(Block-Idle) > 100 counts
(Block-idle) > 100 counts
Stall Setting
(Idle + Block)/2
(Idle + Block)/2