MicroRAE User’s Guide
61
Operation Area and Conditions
Hazardous Areas classified by Zones
PGM-2600 are intended to be used in hazardous areas classified zone 1 or zone 2 within the
temperature range of -20ºC to +60ºC, where gases of explosion groups IIA, IIB or IIC and T4
may be present, and in underground mines endangered by firedamp.
Hazardous Areas classified by Divisions
PGM-2600 is intended to be used in hazardous areas classified for Class I Div. 1 or 2, within the
temperature range of -20º C to +60º C, where gases of explosion groups A, B, C or D and
temperature code T4 may be present.
Instruction For Safe Use
WARNING:
Read and understand instruction manual before operation or servicing.
/AVERTISSEMENT:
Lisez et comprenez le manual d’instructions avant d’utiliser ou
service.
WARNING:
Substitution of components may impact intrinsic safety.
/AVERTISSEMENT:
La substitution de composants peut compromettre la sécurité
intrins
è
que
WARNING:
To prevent ignition of a hazardous atmosphere, batteries must only be
charged in an area known to be non-hazardous. Um = 6.0V. Use only approved charger/
AVERTISSEMENT:
Afin de prevenir l’inflammation d’atmosph
è
res dangereuse, ne
charger le jeu de batteries que dans des emplacement designés non dangereux. Um = 6V
Utilisez uniquement un chargeur approuvé.
Only use approved battery pack: M03-3004-000
Only charge the battery in safe area in the ambient temperature range 0°C ≤ T
amb
≤ 40°C
Use In Hazardous Areas
Equipment intended for use in explosive atmospheres and which has been assessed and certified
according to international regulations may be used only under specified conditions. The
components may not be modified in any way.
The appropriate regulations for service and repair must be properly observed during such
activities.