![Rae microrae Скачать руководство пользователя страница 44](http://html2.mh-extra.com/html/rae/microrae/microrae_user-manual_3372512044.webp)
MicroRAE User’s Guide
44
12. Datalog Transfer, Monitor Configuration, and
Firmware Upgrades Via Computer
Note:
Basic functionality can be adjusted via the MicroRAE’s interface, but more extensive
functionality and parameters can be adjusted using ProRAE Studio II software.
Datalogs can be downloaded from the MicroRAE to a computer, and firmware updates can be
uploaded to the MicroRAE via the USB port on the Travel Charger. Use the included Mini B
USB (5-pin)-to-USB cable to connect the Travel Charger to a computer running ProRAE Studio
II (version 1.10.0 or higher).
12.1.
Downloading Datalogs And Performing PC-Based
Instrument Configuration and Firmware Upgrades
The MicroRAE communicates with a PC running ProRAE Studio II Instrument Configuration
and Data Management software to download datalogs, configure the instrument, or upload new
firmware.
The MicroRAE must be connected to a PC through the supplied Travel Charger and must be in
the PC communications mode.
1.
Use the supplied PC Communications Cable (USB to mini-USB cable) to connect the
Travel Charger to a PC.
2.
Turn on the MicroRAE. Make sure it is running (with the main measurement screen
showing).
3.
Activate the PC communications mode on the MicroRAE by pressing [MODE]
repeatedly, starting from the main measurement screen until you reach the “Comm
Mode” screen.
4.
Press [Y/+]. Measurement and datalogging stop, and the instrument is now ready to
communicate with the PC. The display now says “Ready To PC”.
5.
Start up the ProRAE Studio II software, enter a password, and detect the instrument
following the directions provided in the ProRAE Studio II User’s Guide.
6.
Follow the instructions in the ProRAE Studio II User’s Guide to download the datalog,
configure the instrument settings, or update the MicroRAE’s firmware.
7.
When you are done, press [MODE] to exit the PC communications mode on the MicroRAE.
Note:
After upgrading the MicroRAE's firmware, the instruments will turn off
automatically.
Mini B USB
(5-Pin) to
USB cable
Travel
Charger
Mini B USB
(5-Pin) port
LED
Power
Indicator