OPERATION OF ACCESSORIES
3-1
3. Operation of Accessories
The accessories for the AreaRAE include:
•
Battery charger
•
Alkaline battery adapter
•
External Filter and remote sampling probe
•
Dilution
fitting
•
Calibration adapter
WARNING
To reduce the risk of ignition of hazardous atmospheres,
recharge battery only in area known to be non-hazardous.
Remove and replace battery only in area known to be non-
hazardous. See exception below.
Ne charger les batteries que dans l’emplacement désigné non
dangereux.
Exception: For PGM5x20S, charging battery in a Class I,
Division 2 hazardous location is permitted if installation
complies with 029-STCM-001. The installation requires:
•
Charging circuit to be run within conduit.
•
AreaRAE to be installed within a larger enclosure that
provides conduit entry and requires a tool to open.
•
Charging circuit cable must be terminated with a
Switchcraft S761K or equivalent (see RAE 400-0472-000
for critical requirements) plug with locking collar.
•
The mating charging circuit connectors must be locked
together using the threaded collar before the circuit is
powered.