PROGRAMMING THE AREARAE
4-34
4.7.9. Change Pump Speed
Two speed settings for the pump motor can be selected from
the programming menu: low (default) and high. The “high”
setting (400cc flow per minute) should be used for long lengths
of tubing or when rapid changes in input conditions are
expected. The “Low” setting (300cc flow per minute) may be
selected when the operating conditions are slow to change.
Low may also be used to prolong pump motor life, LEL sensor
life and battery run time.
1. When the “Change Pump Motor Speed?” display appears,
press [Y/+]. The display shows the current pump speed
selection: “Pump Speed = Low?”
2. Press [Y/+] to accept the currently displayed pump speed
and move to the next monitor setup submenu. Press [N/-]
to change to the other speed, “Pump Speed = High?”
Press [MODE] to exit this submenu and move to the next
monitor setup submenu.