OPERATION OF THE AREARAE
2-16
6. The
battery voltage
is the current battery voltage (V)
reading. The reading is updated every second and is
shown with the shutoff voltage:
TOX1
VOC
TOX2
Battery = 7.7V
Shut Off at 6.4V
LEL
OXY
TOX1
VOC
Gamma
Battery = 7.7V
Shut Off at 6.4V
LEL
OXY
AreaRAE
AreaRAE Gamma
Note:
A fully charged battery pack should show 7.7 volts or
higher. When the battery voltage falls below 6.6 volts, a
flashing “Bat” warning message appears. This means that
there is approximately 20 to 30 minutes of operating time
remaining before the battery voltage falls below 6.4 V and
the monitor automatically turns off.
7 through 9 apply to AreaRAE and AreaRAE Steel (non-
Gamma models) only:
7. The
run time reading
is the accumulated time in hours
and minutes the monitor has been on. The reading is
updated in every minute and is displayed with the current
date, time and temperature:
TOX1
VOC
TOX2
Mar 11, ’06
8:30
On = 3:50
20°C
LEL
OXY
8. The
datalog menu
shows the current datalog mode. If the
manual datalog mode is selected, the menu prompts you to
turn datalogging on or off. When “Start Datalog?” displays,
press [Y/+] to turn datalogging on. Likewise, when “Stop
Datalog?” displays, press [Y/+] to turn it off.
9. The monitor displays the selected
LEL
and
VOC gas
name
if the combustible sensor and PID sensor are
installed. When an LEL or VOC gas is selected, the display
features the calculated gas concentration based on the
built-in correction factor for the specified gas.