PROGRAMMING THE AREARAE
4-42
11. “1.00” is the calculated correction factor of the selected gas
in Step 10.
TOX1
VOC
TOX2
Isobutylene
VOC Factor = 1.00?
LEL
OXY
TOX1
VOC
Gamma
Isobutylene
VOC Factor = 1.00?
LEL
OXY
AreaRAE
AreaRAE Gamma
12. If you do not want to modify the VOC correction factor,
press [Y/+] and exit the submenu. To modify this factor,
press [N/-] first. Then starting from the left-most digit, use
[Y/+] or [N/-] to change the digit value, and press [MODE]
momentarily to advance to the next digit. The flashing digit
will move on to next digit to its right. Repeat this process
until all 4 digits of the new factor are entered. Press and
hold [MODE] for 1 second to exit data entry mode. If there
is a change to the existing value, the display shows
“Save?” Press [Y/+] to accept the new value and exit the
gas selection submenu. Press [N/-] to discard the changes.