PROGRAMMING THE AREARAE
4-45
4.8.5 Change PID Lamp Type
This programming menu only applies to the monitor equipped
with a PID detector option. There are two different energy UV
lamps available for the PID sensor: 10.6 eV and 11.7 eV. You
user must select a lamp in order for the instrument to calibrate
correctly.
1. “Change PID Lamp Type?” is the fourth submenu item.
Press [Y/+], and the display shows the current PID lamp
selection: “PID Lamp = 10.6 eV?”
2. Press [Y/+] to accept the currently displayed PID lamp and
automatically advance to Change LEL/VOC Gas Selection.
Press [N/-] to switch to the other lamp. Press [MODE] to
exit this submenu and return to the first submenu.
3. If there is any change to the existing selection, press [Y/+]
and the display shows “Save?” Then, press [Y/+] to accept
or [N/-] to discard and return to the first submenu.
4.8.6 Change RAD Unit (Gamma Models)
This programming menu on the AreaRAE Gamma and
AreaRAE Gamma Steel allows you to choose the display unit
for radiation measurement. The default display unit is µR/h.
1. “Change RAD Unit” is the fifth submenu item. Press [Y/+]
and the current radiation unit selection is: “RAD Unit =
µR/h?”
2.
Press [Y/+] to accept the current selection and go to the
next submenu. Press [N/-] to switch to µSv/h. Press
[MODE] to exit this submenu and return to the first
submenu.
If there is any change to the existing selection, press [Y/+] and
the display will show “Save?” Then, press [Y/+] again to
accept or press [N/-] to discard changes and return to the first
submenu.