261
•
mass
– mass value in current unit is returned,
•
tare
– tare value in calibration/adjustment unit is returned,
•
unit
– determines current (displayed) measuring unit:
Byte no.
Unit
B5 B4 B3 B2 B1
B0 Dec
g gram
0 0 0 0
1
0 1
kg kilogram
0 0 0 0
0
1 2
ct carat
0 0 0
1
0
0 4
lb pound
0 0
1
0
0
0 8
oz ounce
0
1
0 0
0
0 16
N Newton
1
0 0 0
0
0 32
For instance:
Byte no
B5 B4 B3 B2 B1 B0
Value 0 0 0 0 1 0
Scale use kilogram [kg].
•
status
– determines scale’s status
Status
bytes
Task Dec
0
correct measurement (no error message from scale)
1
1 stable
measurement
2
2
scale in zero indication
4
3 scale
tarred
8
4
scale in second measuring range
16
5
scale in third measuring range
32
6
scale reports NULL error
64
7
scale reports LH error
128
8
scale reports FULL error
256
For instance:
byte
no. B8 B7 B6 B5 B4 B3 B2 B1 B0
value
0 0 0 0 1 0 0 1 1