[TI]
Delay
Function determining interval between carrying
out the following dosing process steps. The
function determines pending time for the
following step counted in seconds.
Time
Determination of pending time counted in
seconds
Description
Text description of an interval displayed on
terminal’s touch panel
[Z]
Zero
Zeroing the weighing platform, equal to
operation of the
Æ
0
Å
function key on the
terminal’s front panel
[T]
Tare
Tarring the weighing platform, equal to
operation of the
Æ
T
Å
function key on the
terminal’s front panel
[ST]
Set Tare
Tare setting function, equal to operation of the
<Set tare> function key on the terminal’s front
panel
[CM]
Weight condition
Weight condition function determining time in
which the following step of dosing process on a
weighing platform should be carried out, e.g.:
the following step is carried out if mass (net or
gross) placed on the weighing platform is below
the value set in threshold.
Threshold
Value of mass set in threshold for the condition
Mass
Kind of determined mass for threshold (net or
gross)
Weight condition
Weight condition – “
>=
” or “
<
”
[CI]
Input condition
Conditional function, determining when the
following step should be carried out, depending
on status of terminal’s input. Each input can
take the following status: None – input
disabled; „0” – input with low status; „1” – input
with high status; „/” – input with increasing
tendency (status change from low to high, e.g.:
the moment of pressing a key); „\” – input with
decreasing tendency (status change from high
to low, e.g.: the moment of releasing a key)
[EM]
Give mass
Function activating a so called: “mass in hand”
– ingredient’s mass of a dosing process
supplied in ready packages with precisely
determined mass. The given mass is added to
weighed ingredient’s mass, e.g.: ingredient to
be weighed = 21,8 kg, weighed amount 1,8 kg
placed on the scale, and 20 kg inserted
manually.
130
Содержание WLY
Страница 2: ...2 JULY 2015...
Страница 12: ...Y F1 R series main dimensions Y F1 K series main dimensions 12...
Страница 13: ...Y C2 R series main dimensions Y C2 K series main dimensions Y D2 series main dimensions 13...
Страница 267: ...40 DIAGRAMS OF CONNECTION CABLES Scale computer cable Scale printer CITIZEN EPSON cable Scale Kafka printer cable 267...
Страница 303: ...Print the marked report on a connected printer by pressing Using button view the selected report 303...