37.4.2. Tarring
Syntax:
T CR LF
Possible answers:
T_A CR LF
T_D CR LF
- command accepted and in progress
- command completed
T_A CR LF
T_v CR LF
- command accepted and in progress
- command comprehended but tare range overflow appeared
T_A CR LF
T_E CR LF
- command accepted and in progress
- time limit for stable result exceeded
T_I CR LF
- command comprehended but cannot be executed
37.4.3. Get tare value
Syntax:
OT CR LF
Reply:
OT_TARA CR LF
– command executed
Frame format:
1 2
3
4-12
13 14 15 16
17
18 19
O T space
tare
space
unit
space CR LF
Tare
- 9 characters justified to the right
Unit
- 3 characters justified to the left
Caution:
Tare values are always send in adjustment unit.
37.4.4. Set tare value
Syntax:
UT_TARE CR LF
, where
TARE
– tare value
Possible replies:
UT_OK CR LF
- command completed
UT_I CR LF
- command correct, but not accessible at the moment
ES CR LF
- command incorrect (e.g. incorrect tare format)