14
7.2. Levelling
It is necessary to level the scale, do it by turning its feet. Keep turning the feet
until the air bubble takes central position.
7.3. Start-Up
The weighing device can be connected to the mains only with a power supply
that comes standard with the particular model. Nominal voltage of the power
supply (specified on the power supply data plate) has to be compatible with
the mains nominal voltage.
Procedure:
Plug the power supply to the mains.
Connect the power supply plug to the socket located on the scale's side.
An ON/LOAD diode,
, on indicator housing is turned
on.
Press
key, it is to be found at the top of the operation panel.
Mind to keep the key pressed for about 0.7 s. Operation system and
RADWAG program are loaded, this is signalled with blinking of red
ON/LOAD diode.
Upon completed start-up, the home screen is displayed automatically.
Detailed instruction on how to configure the home screen is
to be found in „PUE 7.1 Indicator Software Manual”.
7.4. Warm-Up Time
For correct scale operation, the workroom temperature must range
b15
C
+30
C.
It takes 30 minutes for the device to warm up.
During the thermal stabilization, the indications on the screen may change.
If you want to carry out adjustment operation, make sure that your device
is thermally stable.