- 33 -
To view the remaining data relating to a respective measurement, first select the measurement in
question, next press
button:
The software automatically shows the data relating to the measurement, data is presented in the
bottom line of the balance display.
To swap between data relating to the measurement, press
or
. The data relating to the
measurement can be printed out by selecting the option <PRINT> and pressing
button.
An exemplary printout:
Date
21.06.2013
Time
13:05:02
User
Product
Tare
0.000 g
Gross
0.000 g
0.000 g
-----------------Cal. report.---------------
Cal. Type
Internal
User
Project
1234567890123459
Date
16.07.2013
Time
13:27:09
Balance ID
10353870
Cal. diff.
-0.004 g
-----------------------------------------------
Signature
…………..........................................
The data that is to be printed out, depends on the settings of parameter P5.3 GLP PRINTOUT.
Parameters selected for the printout (<YES> option), are printed out also on measurement result
printout for WEIGHINGS database (
see sec. 8.2.)
9.5. ALIBI MEMORY
The balance is equipped with “ALIBI”, a type of memory that allows you to save and store up to
100 000 measurements carried out on the balance.
If the “ALIBI” memory is installed in the balance, the saving of the measurements occurs
automatically, by pressing <PRINT> button, without a need of additional actions or settings change.
The additional data is saved along with the measurement.
•
Date of the measurement
•
Time of the measurement
•
Measurement result (mass)
•
Tare value
Содержание AS R2.PLUS
Страница 6: ...6 1 GENERAL INFORMATION 1 1 DIMENSIONS AS PLUS series PS 200 2000 PS 1000 PS 2100 PS 10100...
Страница 97: ...97...
Страница 98: ...98...