- 77 -
CAUTION!
Continuous transmission may be turned on/off by means of a command sent from a computer (go to
COMMUNICATION PROTOCOL section)
13.1.3. Printouts Interval for Continuous Transmission
Parameter <P4.1.3. INTERVAL> enables setting frequency of printouts for continuous transmission.
Frequency of printouts is set in seconds, wherein the accuracy is 0.1s.
The user may set any time value ranging from 1000 to 0.1 s.
The setting is valid for continuous transmission in calibration unit and in current unit activated by
means of the balance (see parameter P4.2.1.2), it is also valid for continuous transmission activated
by command sent from a computer (see point 15 COMMUNICATION PROTOCOL).
13.1.4. Cooperation with E2R
E2R is a control system, it controls all weighing processes performed on a balance. Upon activation
of this option, operations on some Databases are possible only with the use of a computer (options
are not accessible on a balance). In order to start cooperation with E2R system set <YES> value of
<E2R SYSTEM> parameter.
CAUTION!
<E2R System> parameter activation may be carried out by the device manufacturer or authorized persons,
exclusively. Rememeber to switch continuous transmission on to provide correct coperation with the E2R
system.
13.1.5. Printout
Type of the printout sent to the port chosen for the computer.
Procedure:
−
press
button
−
enter <P4 DEVICES> menu
−
enter menu group <P4.1 COMPUTER>
−
enter <P4.1.7 PRINTOUT> parameter
−
select printout:
•
NONE
– no printout selected
•
NSD. PRN.
1÷4 – one of non-standard printouts (
see: non-standard printouts
description)
•
MOB. APP
– special printout (frame with data sent from the balance) recognized by
application used for previewing measurement result on mobile devices.
13.2. PRINTER
Submenu <PRINTER> allows the user to choose a port and a device, to which the data is to be sent
upon pressing
button on a balance. Content of data being transferred is set in parameter
<PRINTOUTS/PRINTOUT GLP>.
Procedure:
-
press
button
-
enter <P4 DEVICES> menu
-
enter menu group: <P4.2 PRINTER>
-
select port to which the printer is to be connected:
COM 1
or
COM 2
– RS 232 port, for connecting a printer
USB 1
type A – USB port, for connecting a PCL printer or EPSON printer
WIFI
– WIFI port for transferring data from balance to a dedicated RADWAG software
(e.g. PW WIN operated on a computer connected to balance via wireless
communication).
Содержание AS R2.PLUS
Страница 6: ...6 1 GENERAL INFORMATION 1 1 DIMENSIONS AS PLUS series PS 200 2000 PS 1000 PS 2100 PS 10100...
Страница 97: ...97...
Страница 98: ...98...