67
10X6-180
For use with Accu-Gold, Accu-Pro, and Accu-Dose, the 10X6-180 is a leakage and low level
measurements chamber. The chamber is designed for leakage measurements with a cross-section of 100
cm (complies with CFR21) and volume of 180 cm . It is also used for very low dose to image receptors.
Chamber
10X6-180
Min Rate
100 nR/s
1 nGy/s
Max Rate
0.6 R/s
4.9 mGy/s
Min Dose
200 nR
2 nGy
Max Dose
2.0 kR
17 Gy
Cine Specifications
N/A
Calibration Accuracy
±4% using X-rays @ 150kVp and 10.2 mm AL HVL
Exposure Rate Dependence
±5%, 20 mR/hr to 2000 R/hr
Energy Dependence
±5%, 30 keV to 1.33 MeV (with buildup material)
Construction
Polycarbonate walls and electrode; conductive graphite
exterior coating; 180 cm active volume; 0.11kg