81
For use with Accu-Gold, the AGMS-D is a Solid-State Dose multi-sensor for Diagnostic Range
Measurements. It is used for single exposure and fluoroscopy. It measures dose, dose-rate, time, kVp with
FFT analysis, Flash HVL, and beam filtration. It features Flash correction of dose for beam quality.
Type
Anode / Filter
Tube
kV
Filt
(um)
hvl (mm
AI) ±10%
or ±0.2
mm AI
whichever
is greater
Rad/Fluoro/Dental
(AGMS-D)
W/AI
General
40-160
±2% (2-24 mm
Al, 40-160 kV)
±2.5% (25-40
mm Al, 50-160
kV)
2-40
mm
1.3-13.5
Rad/Flouro/Dental
(AGMS-D)
Dose using
AGDM
Dose Rate using
AGDM
Dose using
AGDM+
Dose Rate
using AGDM+
Time
150 nGy –
>100 Gy,
±5% (17 R
– >11 kR,
±5%)
150 nGy/s – 350
mGy/s, ±5% (1.0
mR/min – 2.4
kR/min, ±5%)
80 nGy - >100
Gy, ±5% (9.1
uR - >11 kR,
±5%)
80 nGy/s - 150
mGy/s
±5% (0.55
mR/min - 1.0
kR/min, ±5%)
1 ms – 300 s
± 0.1% + 0.2
ms
Reference condition: 12
°
Tungsten anode