DDX6-WL
For use with Accu-Gold, Accu-Pro, and Accu-Dose, the DDX6-WL is a solid-state dose sensor for
diagnostic range measurements. It is a small solid-state sensor for single exposures and fluoroscopy in
the Diagnostic range. It measures dose, dose-rate, and time and is passively corrected for beam quality.
It may be used simultaneously with an AGMS-D multi-sensor in entrance / exit exposure measurements.
It is especially recommended for low dose-rate measurements, e.g. pediatric dose-rates.
Sensor
DDX6-WL
Min Rate
50 nR/s
500 pGy/s
Max Rate
0.65 R/s
5.7 mGy/s
Min Dose
200 nR
2 nGy
Max Dose
1.8 kR
15 Gy
Accuracy
±5% using X-rays @ 80kVp with 2.5 mm Al total filtration
(IEC 61267 RQR-6)
Energy Dependence
+5% from 50 kVp to 120 kVp at 2.5 mm Al
Filtration Dependence
+5% to -10% from 2.5 mm Al to 23 mm Al