95
Dose / Total Dose
yes
no
Ratio of the measurements of extracted dose to total dose.
Dose Rate
yes
yes
Dose Rate
Dose Ratio AGMS / DD
yes
yes
Multi-sensor dose divided by dose-diode dose.
Dose Ratio AGMS / IC
yes
yes
Multi-sensor dose divided by ion-chamber dose
End Average kV
Yes
Yes
END is useful for capturing the radiation values at the end
of an exposure. For exposures longer than 100 ms, and
less than 1 second, END captures the last 15% of the
exposure. For exposures longer than 1 second, END
captures the last 1 second. End ignores the he last 5 ms
when calculating the values.
END results are available for:
AGMS (avg. kV, filtration, HVL, dose rate)
IC (dose rate)
D (dose rate)
End Dose Rate
Yes
Yes
End Filtration
Yes
Yes
End HVL
Yes
Yes
Filter yes
yes
Tube filter setting used for this measurement.
Filtration yes
yes
Measured beam-filtration thickness.
HVL yes
yes
Half value layer result.
kV Duration
yes
no
Width of the kV signal
kV Peak Ripple Frequency
yes
no
The frequency of the largest ripple component in the
spectrum from the Fourier Transform.
kV Pulse Repetition
Frequency
yes no
Pulse-repetition frequency as calculated from the kV wave
in the region of interest with synchronization to pulse edges
when possible.
kV RMS Noise
yes
no
Root-mean-square of the noise and ripple in the kV values
above the adaptive threshold used in the calculation of kVp.
kV Zero to Peak Ripple
yes
no
Zero-to-peak amplitude of the largest ripple component in
the spectrum from the Fourier Transform.
kVp yes
yes
Average peak voltage
PPV yes
no
The Practical (PPV) mode computes a weighted average
of the kV waveform in the region of interest, with weighting
toward higher-kV values, employing the algorithm of
Kramer, Selbach and Iles.
Total Dose
yes
yes
Total dose of source measurement.
Ion Chamber
Dose yes
yes
Dose Integral
Dose / mAs
yes
no
Dose / mAs
Dose / Pulse
yes
yes
Dose / Pulse
Dose / Total Dose
yes
yes
Ratio of the measurements of extracted dose to total dose.
Dose Rate
yes
yes
Dose Rate
Dose Ratio IC / AGMS
yes
yes
Dose Ratio IC / AGMS
Dose Ratio IC / DD
yes
yes
Dose Ratio IC / DD
End Dose Rate
Yes
Yes
(see END for AGMS)
Temperature yes
yes
Temperature measured by ion chamber.
Total Dose
yes
yes
Total dose of source measurement.
Dose Diode
Dose yes
yes
Dose Integral
Dose / mAs
yes
no
Dose / mAs
Dose / Pulse
yes
yes
Dose / Pulse
Содержание ACCU-GOLD2
Страница 1: ...dddddddd ACCU GOLD2 USER GUIDE For use with all Accu Gold digitizers and Accu Gold Touch Pro...
Страница 34: ...33...
Страница 40: ...39 Daughter designation Extracted Region...
Страница 43: ...42...
Страница 47: ...46...
Страница 59: ...58...
Страница 99: ...98 Appendix G Carrying cases Accu Gold Soft Cases Model AGSC...
Страница 100: ...99 Hard Cases 40V94R Accu Gold Carry Case 2 layer 5 125 mm deep 40V94S Single Layer Carry Case 4 100 mm deep...
Страница 105: ...104 Test Stands and Positioners ADTS W Accu Dose Test Stand AGTS W Accu Gold Test Stand...