72
10X6-0.18
For use with Accu-Gold, Accu-Pro, and Accu-Dose, the 10X6-0.18 is a high dose rate chamber used for
in-beam measurements of high-intensity gamma radiation. The fully guarded chamber is mounted at the
end of a 3-m low-noise triax cable.
Chamber
10X6-0.18
Min Rate
50 R/s
500 nGy/s
Max Rate
720 R/s
6.31 Gy/s
Min Dose
200 R
2 Gy
Max Dose
2 MR
17 kGy
Cine Specifications
N/A
Calibration Accuracy
±4% @ 60 Co
Exposure Rate Dependence
±2%, 3 mR/s to 720 R/s
Energy Dependence
±5%, 45 keV to 1.33 MeV
Active Length
Construction
8.1 mm ± 1 mm
C552 air-equivalent material walls and electrode; polyacetal
exterior cap, 0.18 cm active volume, 3-m triax cable
Calibrations at
60
Co and
137
Cs are available.
Содержание ACCU-GOLD2
Страница 1: ...dddddddd ACCU GOLD2 USER GUIDE For use with all Accu Gold digitizers and Accu Gold Touch Pro...
Страница 34: ...33...
Страница 40: ...39 Daughter designation Extracted Region...
Страница 43: ...42...
Страница 47: ...46...
Страница 59: ...58...
Страница 99: ...98 Appendix G Carrying cases Accu Gold Soft Cases Model AGSC...
Страница 100: ...99 Hard Cases 40V94R Accu Gold Carry Case 2 layer 5 125 mm deep 40V94S Single Layer Carry Case 4 100 mm deep...
Страница 105: ...104 Test Stands and Positioners ADTS W Accu Dose Test Stand AGTS W Accu Gold Test Stand...