1261B User Manual
Racal Instruments
Ó
1995
Using The Enhanced Monitoring System 4-9
Programming The
VXI Slot
Temperature
Tolerance Limits
Each VXI slot is individually monitored to ensure that the
temperature
does not exceed its programmable limit.
SENSE:TEMPerature<N> Commands
These commands set either the absolute temperature limits or
relative temperature limits for the VXI chassis. The <N>
following the TEMPerature indicates which temperature is being
monitored
.
TEMP1:RANG:UPP <temperature> Temperature of VXI slot 0, relative
TEMP2:RANG:UPP <temperature> Temperature of VXI slot 1, relative
TEMP3:RANG:UPP <temperature> Temperature of VXI slot 2, relative
TEMP4:RANG:UPP <temperature> Temperature of VXI slot 3, relative
TEMP5:RANG:UPP <temperature> Temperature of VXI slot 4, relative
TEMP6:RANG:UPP <temperature> Temperature of VXI slot 5, relative
TEMP7:RANG:UPP <temperature> Temperature of VXI slot 6, relative
TEMP8:RANG:UPP <temperature> Temperature of VXI slot 7, relative
TEMP9:RANG:UPP <temperature> Temperature of VXI slot 8, relative
TEMP10:RANG:UPP <temperature> Temperature of VXI slot 9, relative
TEMP11:RANG:UPP <temperature> Temperature of VXI slot 10, relative
TEMP12:RANG:UPP <temperature> Temperature of VXI slot 11, relative
TEMP13:RANG:UPP <temperature> Temperature of VXI slot 12, relative
TEMP14:RANG:UPP <temperature> Ambient Temperature
TEMP15:RANG:UPP <temperature> Temperature of VXI slot 0, absolute
TEMP16:RANG:UPP <temperature> Temperature of VXI slot 1, absolute
TEMP17:RANG:UPP <temperature> Temperature of VXI slot 2, absolute
TEMP18:RANG:UPP <temperature> Temperature of VXI slot 3, absolute
TEMP19:RANG:UPP <temperature> Temperature of VXI slot 4, absolute
TEMP20:RANG:UPP <temperature> Temperature of VXI slot 5, absolute
TEMP21:RANG:UPP <temperature> Temperature of VXI slot 6, absolute
TEMP22:RANG:UPP <temperature> Temperature of VXI slot 7, absolute
TEMP23:RANG:UPP <temperature> Temperature of VXI slot 8, absolute
TEMP24:RANG:UPP <temperature> Temperature of VXI slot 9, absolute
TEMP25:RANG:UPP <temperature> Temperature of VXI slot 10,absolute
TEMP26:RANG:UPP <temperature> Temperature of VXI slot 11,absolute
TEMP27:RANG:UPP <temperature> Temperature of VXI slot 12,absolute
<temperature> range:
0.0 to 140.0 for relative temperature (default at 30.0 degrees C)
0.0 to 75.0 for absolute temperature (default at 55.0 degrees C)
Example: “
SENS:TEMP22:RANG:UPP 58
” set the alarm
temperature for slot 7 to 58 degrees C.
T
hese commands queries either the absolute temperature limits
or relative temperature limits of the monitored VXI slot. The <N>
following the TEMPerature indicates which temperature slot is
being queried.