1261B User Manual
SCPI Command Reference 5-10
Racal Instruments
Ó
1995
Command
Parameter Limits
The parameter limits for the numeric commands are shown in
Table 5-2.
Table 5-2, Defaults, Minima and Maxima
Command Default
Minimum
Maximum
Notes
SENSE:VOLT1:RANGE:UPPER 5.4
5.0
40.0
+5V PS voltage upper tolerance
SENSE:VOLT1:RANGE:LOWER
4.6
0.0
5.0
+5V PS voltage lower tolerance
SENSE:VOLT2:RANGE:UPPER
-4.8 -5.2
0.0
-5.2V
PS voltage upper tolerance
SENSE:VOLT2:RANGE:LOWER
-5.6
-40.0
-5.2
-5.2V PS voltage lower tolerance
SENSE:VOLT3:RANGE:UPPER -1.8
-2.0 0.0
-2V PS voltage upper tolerance
SENSE:VOLT3:RANGE:LOWER -2.2
-16.0 -2.0
-2V PS voltage lower tolerance
SENSE:VOLT4:RANGE:UPPER 25.9 24.0
100.0
+24V PS voltage upper tolerance
SENSE:VOLT4:RANGE:LOWER 22.1
0.0 24.0
+24V PS voltage lower tolerance
SENSE:VOLT5:RANGE:UPPER
-22.1 -24.0
0.0
-24V
PS voltage upper tolerance
SENSE:VOLT5:RANGE:LOWER
-25.9
-100.0
-24.0
-24V PS voltage lower tolerance
SENSE:VOLT6:RANGE:UPPER
12.9 12.0
100.0
+12V PS voltage upper tolerance
SENSE:VOLT6:RANGE:LOWER
11.1
0.0
12.0
+12V PS voltage lower tolerance
SENSE:VOLT7:RANGE:UPPER
-11.1 -12.0
0.0
-12V
PS voltage upper tolerance
SENSE:VOLT7:RANGE:LOWER
-12.9
-100.0
-12.0
-12V PS voltage lower tolerance
SENSE:CURR1:RANGE:UPPER 85.6 0.0
100.0
+5V
PS current upper tolerance
SENSE:CURR2:RANGE:UPPER 64.2
0.0
75.0
-5.2V PS current upper tolerance
SENSE:CURR3:RANGE:UPPER 32.1
0.0
37.5
-2V PS current upper tolerance
SENSE:CURR4:RANGE:UPPER 12.9 0.0
15.0
+24V PS current upper tolerance
SENSE:CURR5:RANGE:UPPER 12.9 0.0
15.0
-24V PS current upper tolerance
SENSE:CURR6:RANGE:UPPER 13.9 0.0
16.3
+12V PS current upper tolerance
SENSE:CURR7:RANGE:UPPER 13.9 0.0
16.3
-12V PS current upper tolerance
SENSE:FREQ:RANGE:UPPER
5200.0 2000.0
7650.0
All
Fans, upper RPM tolerance
SENSE:FREQ:RANGE:LOWER
2000.0
500.0
7650.0
All Fans, lower RPM tolerance
SENSE:TEMP14:RANGE:UPPER
55.0
0.0
140.0
Ambient temperature upper tol
SENSE:TEMP<X>:RANGE:UPPER 30.0
0.0 140.0
VXI slot temp rise tol.
<X> = 1 to 13
SENSE:TEMP<X>:RANGE:UPPER 55.0
0.0 140.0
VXI slot temp Absolute.
<X> = 15 to 27
SENSE:TIME1:RANGE:UPPER
31536000
(365 days)
0.0
125 years
Power-On elapsed time limit
SENSE:TIME2:RANGE:UPPER
157680000
(5 x 365 days)
0.0
125 years
Cumulative power-on time limit
SENSE:TIME3:RANGE:UPPER
15552000
(180 days)
0.0
125 years
Filter service power-on time limit
SENSE:VXI:IACK<X>:LIMIT
0
0
255
IACKn limit (<X> = 1 to 7)
SENSE:VXI:BERR:LIMIT
0
0
255
VXI Bus Error limit
VXI:CONF:MON:TRIG:DELAY:TIME 0.0
0.0
1048575968.75
(nanoseconds)
Trigger Delay time
resolution = 31.25 nanosecond