1261B User Manual
Racal Instruments
Ó
1995
Operating The 1261B 3-15
Basic Functional
Check
The functional check consists of checking the 1261B power supply
voltages to the VXIbus limits at rear panel connector J415 using a
digital voltmeter.
Referring to
Table 3-10
, connect one lead of the voltmeter to a
supply pin at J415. Connect the reference lead of the voltmeter to
one of the ground pins. Compare each voltage reading to the
values listed in
Tables 3-10.
Table 3-9, Power Supply Voltages at J415
J415 Pin
Supply
VXI Acceptable
Voltage Range
11
+5V
4.875V to 5.25V
15
-12V
-12.6 to 11.64V
1
-24V
-25.2V to -23.28V
3
-2V
-2.1V to -1.9V
17
+12V
11.64V to 12.6V
16
+24V
23.28V to 25.2V
8
-5.2V
-5.46V to -5.044V
2,4,14 or 24
Logic Ground
If the voltages are within the specified ranges, the mainframe
complies with VXI voltage limit specifications.