1261B User Manual
Configuring The 1261B 2-2
Racal Instruments
Ó
1995
1261B (bench or rack mount) and follow the appropriate
installation instructions
.
Installing VXI
Modules
Install C-size modules directly into the mainframe by first placing
the module’s card edges into the front module guides (top and
bottom). Slide the module to the rear of the mainframe until the
module connectors mate solidly with the backplane connectors.
Secure the module’s front panel to the mainframe using the
module’s front panel mounting screws.
The easiest way to install B-size modules is with a “B to C-size
adapter” allowing a B-size module to be installed just like a C-size.
Installation and
Removal of Card
Guide Covers
(Airflow
Directors)
In order to improve cooling of used slots in the VXI mainframe, a
limited number of airflow directors, P/N 456271, may be installed
at empty slot locations to redirect otherwise wasted airflow.
CAUTION
To maintain a balanced airflow throughout the
system when using airflow directors, do not cover
more than six empty slots at any time. If module
density is low, space airflow directors evenly across
the mainframe.
Mainframe Rear
Mainframe Front
Figure 2-1, Airflow Directors
1.
Facing the front of the VXI mainframe, with one hand hold
the airflow director by the front end where the “Racal” logo
appears.
2.
Select the slot to be covered by the director, and place the
hook at the rear of the director into the rectangular hole at
the rear of the cardguide (1). Slight downward pressure on
top of the director (with the other hand) at the rear may be
required to engage the hook into the cardguide.