![QwikCheck Gold Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/qwikcheck/gold/gold_user-manual_3810243032.webp)
QwikCheck™
Sperm Quality Analyzer Guide Version 1.00 I-Button WHO 5th 06_AUG_2012
32
Self Test Parameters:
Fill-in the QwikCheck SELF TEST PARAMETERS from the SELF TEST printout in V-Sperm:
From the
MAIN MENU
go to:
SERVICE>PRINT SELF TEST DATA AND DEFAULT
SETTINGS>SELF TEST DATA.
Verify that the parameters listed below fall within the established range
Highlight the discrepancies and report to MES
Parameter
S/W Ver. 2.48
Criteria
SYSTEM Self-Test
Parameters
Ref. 1
150 – 350 mV
LED Current 1
5 – 20 mA
Original
value
Amplitude
50 – 100 mV
Count (#12)
26 – 36
Zero Level
500 – 525
Ref. 2
2500 – 3500
LED Current 2
10 – 32 mA
Original
value
TSC 1 or CONC 1 0 – 1 M/ml
TSC 2 or CONC 2 50 – 150 M/ml
TSC 3 or CONC 3 300 – 600 M/ml