Customer Support/Troubleshooting
: To quickly assess if the QwikCheck GOLD is
in good calibration, refer to the table and values below. The S/W Ver. 1.00 column
provides reference ranges for calibration. Enter information from the SELF TEST DATA
in the “QwikCheck Value" column to see if the system requires troubleshooting.
#
Parameter
S/W Ver. 1.00
QwikCheck
Value
Pass
Fail
1.
Ref 1
150 – 350 mV
2.
LED Cur 1
5 – 25 mA
3.
Amplitude
50 – 100 mV
4.
Zero Level
500 - 525
5.
Ref 2
2500 – 3500 mV
6.
LED Cur 2
10 – 32 mA
7.
CONC. 1
0 – 1 M/ml
8.
CONC. 2
50-150 M/ml
9.
CONC. 3
300-600 M/ml
10.
Count (Service
Data, Item #12)
26 - 36
I-Button
loading
SECTION 6: Add I-Button Tests
To add I-button tests, go to:
MAIN MENU> ADD I-BUTTON TESTS
and
press
ENTER.
The screen below will be displayed.
In each box of 50 testing capillaries, a new i-button is included that has 50
test credits loaded on the button.
Follow the on-screen instructions, holding the new i-button firmly against the
i-button port located on the side of the QwikCheck GOLD system.
The screen above will be displayed when the i-button is successfully loaded!
MAIN MENU
TEST NEW PATIENT
RUN CONTROLS
ADD I-BUTTON TESTS
SERVICE
TO ADD MORE TESTS
HOLD NEW I-BUTTON AGAINST PORT
AND PRESS ENTER
PRESS ESC TO EXIT
# TESTS ADDED: 50
# OF TESTS NOW REMAINING: 76
PRESS ESC TO EXIT