![QwikCheck Gold Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/qwikcheck/gold/gold_user-manual_3810243012.webp)
QwikCheck™
Sperm Quality Analyzer User Guide Version 1.00 I-Button WHO 5th _06_AUG_2012
12
Low Quality
Test Results
FROZEN,
WASHED
SAMPLE
TESTING
When a sample is LOW QUALITY, test results may be reported as < or > when
parameters fall below the dynamic range of the system. Only the following parameters
will be reported: Sperm Concentration, Motility, SMI and Motile Sperm Concentration due
to the limited number of cells, very low motility and/or poor morphology.
Examples of test results reported in this manner are seen in the screens below:
Run FROZEN and WASHED samples in the same manner by selecting SAMPLE TYPE:
FROZEN (or WASHED) from the MAIN MENU > TEST NEW PATIENT > ENTER SAMPLE
DATA screen. Follow the on-screen instructions.
FROZEN samples require only 20 microliters of sample and motility parameters
only are reported
WASHED sample can be run with a large or small volume sample (motility
parameters only will be report if using 20 microliters)
Printing
If the QwikCheck was set to automatically print test results on the label maker
they will now print.
If the default was not set, press the PRINT button on the keypad.
TEST RESULTS
SPERM CONC. <2.0 M/ml
TOTAL MOTILITY <PR+NP> < 22 %
PROG. MOTILITY <PR> %
NONPROG. MOTILITY <NP> %
IMMOTILITY <IM>
%
MORPH. NORM. FORMS, WHO 5TH
%
TEST RESULTS
MSC < 0.2 M/ml FSC M/ml
PMSC M/ml VELOCITY mic/sec
SMI 17
TOTALS PER EJACULATE
SPERM # N.A. MOTILE SPERM N.A.
PROG.SPERM N.A. FUNC SPERM N.A.
MORPH. NORM. SPERM N.A.
TESTING
LOW QUALITY SAMPLE
TESTING WILL TAKE 2 MORE MINUTES