APPENDIX 9: Assayed Control – QwikCheck-beads™ 

 

A QUALITY CONTROL MATERIAL FOR AUTOMATED AND MANUAL SPERM COUNTING SYSTEMS 

Introduction and Intended Use: 

Q

wik

C

heck

TM 

beads

 

is  an  in-vitro  use  only  external  quality  control  material  for  automated  and  manual  sperm  counting  systems.  It  has  been 

developed as a tool to assess the accuracy and precision of the l

aboratory’s sperm counting methods by providing a known target value and +/

range.    The  beads  were  developed  for  use  on  the  SQA-V  semen  analyzer  however,  they  are  also  labeled  for  manual  proficiency  testing  and 
calibration on hemacytometers such as Neubauer counting chambers, Makler chambers and conventional fixed coverslip chambers.   
 

Q

wik

C

heck

TM 

beads

 

is supplied in a kit containing known concentrations of 4-micron latex beads suspended in an aqueous solvent and negative 

concentration/motility control. The beads should be run according to the directions below for the type of system used in the lab. According to the 

CLIA  ’88  regulations,  “…for  most  moderately  complex  tests,  the  general  requirement  is  to  analyze  two  levels  of  QC  m

aterials  on  each  day  of 

testing.”    It  is  recommended  that  QwikCheckTM  beads  be  run  on  the  SQA

-V  automated  and  visualization  systems  prior  to  each  day  of  semen 

analysis testing. 

 

For in-vitro use only: 

Each kit contains two known concentrations of 

Q

wik

C

heck

TM 

beads

 

in two 5 ml aliquots and one 5 ml negative concentration and motility control.  

Store the beads at room temperature (20-25ºC or 65-77ºF).  The expiration date assumes that 

Q

wik

C

heck

TM 

beads

 

are stored at room temperature 

in their original containers and tightly capped to prevent evaporation. Open vial shelf-life is 90 days (tightly capped, stored in the original container at 
room temperature or refrigerated @ 5-15 degrees Centigrade).   

 

Target Value and +/- Ranges 

 

Target Values and +/- Ranges unique to the batch have established for each batch of QwikCheck beads. Each box and each control bottle is 
labeled with these Target Values and Ranges.  In addition, the MES website: 

www.mes-global.com

  provides Batch Release Forms by batch # with 

details.  

 

Warning:  

Contains 0.1% Sodium Azide as a preservative. Other ingredients are not harmful due to the low concentration in the beads.  For additional 
information, please refer to the QwikCheck-beads Material Safety Data Sheet #

 

QCB 001

 

 

Basic Instructions for using 

Q

wik

C

heck

TM 

beads

 

 

1. 

Mix the beads in the closed container before use to evenly distribute the beads in the suspension and insure accurate testing outcomes. 
To mix, gently rotate the beads by hand (do not use a vortex).   

2. 

Open the container and immediately withdraw a sample of the control material. Tightly close the container after withdrawing the sample. 

3. 

Follow the detailed instructions below for the type of system/chamber used.  

 

Instructions for running QwikCheck

TM 

beads QC material on automated and manual sperm counting systems:   

 
SQA-V Automated System:

 

1. 

Refer to the SQA-

V User Guide “CONTROLS” section for an explanation of how to set

-up the SQA-V to test automated Level 1/Level 2 

and Negative Control QwikCheck

TM 

beads.  Follow the SQA-

V onscreen instructions in the “Controls” section of the SQA

-V User Guide:  

2. 

Gently rotate the closed container of beads by hand (do not use a vortex).  

3. 

Aspirate a sample of the beads

 

or negative control into the SQA-V capillary in the same manner you would fill the capillary for a normal 

volume specimen.  Make sure that the cuvette section of the SQA-V capillary is completely full of liquid and free of bubbles.  

4. 

Insert the testing capillary into the SQA-V measurement chamber when prompted.  

5. 

Transfer the Control test results to V-Sperm when prompted. 

 

SQA-V Visualization System using a standard slide:

 

1. 

Refer to the SQA-

V User Guide “Operating the Visualization System” to understand how to use a standard slide in the SQA

-V.  

2. 

Gently rotate the closed container of beads by hand (do not use a vortex). 

3. 

Pipette 10 µL of QwikCheck

TM 

beads onto a standard slide, cover with a 22x22 mm coverslip to provide a 20-micron sample depth. 

4. 

If air bubbles or liquid spillage occurs, prepare a new slide to ensure an accurate reading. 

5. 

Insert  the  slide  into  the  slide  adaptor  and  then  into  the  visualization  chamber  of  the  SQA-V.  Press  Zoom-Out  all  the  way  to  set  the 

magnification at x300 and FREEZE the image using V-Sperm software on the computer. 

6. 

Count the beads manually per WHO guidelines:  Duplicate counts of at least 200 beads are required

 

(turn the slide adaptor knob to view 

multiple fields). Divide the final number of beads counted in the multiple fields by the number of screens viewed. Each bead on the SQA-V 
screen (V-Sperm III or the SQA-V screen) represents 1 M/ml.  

7. 

Refer to WHO Manual 5

th

 Edition table 2.2 to determine if the duplicate counts are acceptable. 

 

Содержание Gold

Страница 1: ...er User Guide Version 1 00 I Button WHO 5th _06_AUG_2012 1 QwikCheck S p e r m Q u a l i t y A n a l y z e r U S E R G U I D E W H O 5 t h V e r s i o n V e r s i o n 1 0 0 I B u t t o n Catalog F A 0...

Страница 2: ...ing 12 SECTION 5 Controls and QC Control Set Up and Testing 13 Set Up Assayed Control 13 Set Up Non Assayed Control 13 Running CONTROLS 14 Electronic Self Test and Auto Calibration 15 SECTION 6 I Butt...

Страница 3: ...DIX 7 Dilution Media QwikCheck Dilution 28 APPENDIX 8 Treating Viscous Samples QwikCheck Liquefaction 29 APPENDIX 9 Assayed Control QwikCheck beads 30 APPENDIX 10 Service Report 31 APPENDIX 11 QwikChe...

Страница 4: ...d QwikCheck Testing Capillary Keypad Navigation Use NUMERIC keys to enter data ARROW keys to move to the next field Press ENTER to select menu options confirm data entries and to move to the next scre...

Страница 5: ...OT SPERM M Immotility IMMOTILTIY IM Progressively Motile Sperm Concentration PMSC M ml Sperm Morphology normal forms MORPH NORM FORMS WHO 5th Total Progressively Motile Sperm ejaculate PROG SPERM M Mo...

Страница 6: ...ed on proprietary algorithms Step 3 Motility Tens of thousands of sperm cells are analyzed in the thin section of the testing capillary as they move through a light beam in the system causing light di...

Страница 7: ...key on the keypad and auto calibration will begin This process takes 5 7 minutes When the system stabilization and auto calibration processes are complete a series of tests will be run Do not insert...

Страница 8: ...perm product labeling Select the type of controls to be run Latex Beads Stabilized Sperm Select the LEVEL 1 2 or NEGATIVE Enter the LOT EXPIRATION DATE and TARGET VALUE from the product labeling The Q...

Страница 9: ...Low volume samples diluted 1 1 with QwikCheck dilution media can be used according to User Guide instructions WASHED Sample enriched or prepared for artificial insemination using a commercial media to...

Страница 10: ...y with a Normal Volume Sample Insert the testing capillary into the measurement chamber of the QwikCheck GOLD when instructed by the screen below Testing will begin automatically do not touch the syst...

Страница 11: ...report accurate results only if the sample has been diluted 1 1 prcisely and the sample is adequate to fill the testing capillary after dilution To run a LOW VOLUME sample select option 2 from the scr...

Страница 12: ...N MENU TEST NEW PATIENT ENTER SAMPLE DATA screen Follow the on screen instructions FROZEN samples require only 20 microliters of sample and motility parameters only are reported WASHED sample can be r...

Страница 13: ...set up updated To do this Go to MAIN MENU SERVICE SET UP CONTROLS to view the screen below Select the type of control LATEX BEADS or STABILIZED SPERM Select the level of the control LEVEL 1 LEVEL 2 or...

Страница 14: ...MAIN MENU RUN CONTROLS and press ENTER When the screen below is displayed Select the CONTROL LEVEL 1 2 or NEGATIVE LEVEL 3 that is being tested Press ENTER to continue Fill the testing capillary with...

Страница 15: ...k will report failures see section on error and warning messages and freeze the system if the system is not within the established self test ranges Prior to testing a sample Auto calibration verificat...

Страница 16: ...1 0 1 M ml 8 CONC 2 50 150 M ml 9 CONC 3 300 600 M ml 10 Count Service Data Item 12 26 36 I Button loading SECTION 6 Add I Button Tests To add I button tests go to MAIN MENU ADD I BUTTON TESTS and pre...

Страница 17: ...S Service Personnel A code is required to access SERVICE PERSONNEL This option allows a qualified service technician to access calibration and maintenance settings Print system Default Settings The sy...

Страница 18: ...ion Call technical support if failure recurs Self test Failed Ensure there is no testing capillary in the measurement compartment Remove the QwikCheck from sources of electronic noise cell phones etc...

Страница 19: ...rt if this message is displayed again Prepare for technical support by printing a copy of the SERVICE DATA Press the SERVICE key on the keypad to activate the SERVICE MENU screen Select PRINT SELF TES...

Страница 20: ...elow the piston head and pull the piston back slowly while keeping the tip of the capillary well below the sample level and below any surface bubbles Figure 1 Continue to aspirate the sample until it...

Страница 21: ...Fill only the thin capillary chamber with 20 micro liters of semen Figure 1 3 The quantity aspirated can be determined by the gradations on the 1 ml syringe Aspirate the sample until it just appears...

Страница 22: ...ibrous material capillary fig 1 Moisten with ONE drop of cleaning fluid shaking off excess fluid Insert into the measurement compartment fibrous material facing up Move back and forth a few times Repe...

Страница 23: ...M FORMS WHO 5th 4 WHO 5th manual Motile Sperm Concentration MSC 6 M ml MES Progressively Motile Sperm Concentration PMSC 5 M ml MES Functional Sperm Concentration FSC Velocity Average path velocity VA...

Страница 24: ...s of concentration In accordance with CLIA regulations such a control is used to demonstrate operator proficiency using the microscope and for validation of automated sperm counting methods The latex...

Страница 25: ...The coefficients of variation CV characterizing precision were calculated for Sperm Concentration and Motility and were below 6 Specificity To achieve analytical specificity a specific wave length of...

Страница 26: ...ntrifuged at 600g for 15 minutes The seminal plasma was decanted and the pellets were re suspended in washing media DPBS HepesHTF Sequential dilutions were run in 4 SQA systems Limitations of method D...

Страница 27: ...by bacteria trichomonads or erythrocytes present in the specimen QwikCheck test strips were evaluated by Medical Electronic Systems Ltd MES for use as a qualitative indicator WBC s 1M ml of WBC s in h...

Страница 28: ...onged exposure to light Do not use if the solution contains precipitate or is cloudy INSTRUCTIONS FOR USE AUTOMATED SQA V 1 Measure the volume of the neat semen sample 2 If the volume is less than 0 5...

Страница 29: ...e stored at room temperature INSTRUCTIONS FOR USE 1 Select one vial of Chymotrypsin 2 Tap the vial to move the contents to the bottom of the vial prior to opening 3 Add the entire contents of one vial...

Страница 30: ...ion please refer to the QwikCheck beads Material Safety Data Sheet QCB 001 Basic Instructions for using QwikCheckTM beads 1 Mix the beads in the closed container before use to evenly distribute the be...

Страница 31: ...he QwikCheck GOLD Contact your local distributor for the initial calibration parameters These parameters should not have changed Parameter Service Report Item User Report Initial Calibration settings...

Страница 32: ...LT SETTINGS SELF TEST DATA Verify that the parameters listed below fall within the established range Highlight the discrepancies and report to MES Parameter S W Ver 2 48 Criteria SYSTEM Self Test Para...

Страница 33: ...TA REF1 230 REF2 2925 LED1 9 LED2 24 AMP 65 CONC1 0 0 SMI 409 CONC2 104 6 ZL 508 CONC3 418 9 AW 14987 SERVICE DATA 1 16 7 0 000 13 100 2 5 8 115 14 100 3 150 9 10 15 1 70 4 28 10 6 16 100 5 0 95 11 13...

Страница 34: ...ly run semen samples based on that standard SQA V Set Up From the MAIN MENU select SERVICE SET UP SYSTEM DEFAULTS to view the screen below Select a CONC concentration STD standard 1 or 2 based on the...

Отзывы: