SECTION 8: Error Messages and Warning Messages
Stabilization Failed:
Ensure there is no testing capillary in the measurement compartment.
Remove the SYSTEM from sources of electronic noise (cell phones, etc.) and
vibrations (centrifuge).
Clean measurement compartment (refer to Appendix).
Reboot the QwikCheck system without a testing capillary in the chamber:
Turn system
OFF
then back
ON
at the main switch on the rear panel.
Press the front panel
ON/OFF
key to begin Auto-Calibration/Stabilization.
Call technical support if failure recurs.
Self-test Failed:
Ensure there is no testing capillary in the measurement compartment.
Remove the QwikCheck from sources of electronic noise (cell phones, etc.)
and vibrations (centrifuge).
Clean measurement compartment (refer to Appendix).
Reboot the QwikCheck system without a testing capillary in the chamber:
Turn the system
OFF
then back
ON
at the main switch on the rear panel.
Press the front panel
ON/OFF
key to begin Auto-Calibration and
Stabilization.
Call technical support if this message is displayed again. Prepare for technical
support by printing a copy of the
SERVICE DATA:
Press the SERVICE key on the keypad to activate the
SERVICE MENU
screen.
Select:
PRINT SELF TEST DATA AND DEFAULT SETTINGS>SELF
TEST DATA.
Press
ENTER