Electronic Noise:
Ensure there is no testing capillary in the measurement compartment.
Remove the QwikCheck from sources of electronic noise (cell phones, etc.)
and vibrations (centrifuge).
Clean measurement compartment (refer to Appendix) and after cleaning:
Turn the system
OFF
then back
ON
at the main switch on the rear panel.
Press the front panel
ON/OFF
key to begin Auto-Calibration and
Stabilization.
From
MAIN
menu: Select
TEST NEW PATIENT
and rerun the test.
Call technical support if this message is displayed again. Prepare for technical
support by printing a copy of the
SERVICE DATA
:
Press the SERVICE key on the keypad to activate the
SERVICE MENU
screen.
Select:
PRINT SELF TEST DATA AND DEFAULT SETTINGS>SELF
TEST DATA.
Press:
ENTER
Concentration Out of Range
Testing Semen Sample:
A message will appear indicating that the tests results for Sperm Conc and/or
MSC are beyond the upper limits of the dynamic range established by the
manufacturer for testing. This message will appear when:
SPERM CONC > 500 M/ml or
MSC > 450 M/ml
Review sample handling technique (see Appendix "Filling the SQA Capillary").
Re-test the sample using a new SQA capillary. If the message appears again,
reboot the system.
Call for technical assistance if problem persists.