Scalar i6000 Installation Guide
339
Calibrating the Digital Level
This section describes how to calibrate the Pro 3600 digital level.This procedure re-calibrates the level
through its entire 360 degree range by recording four horizontal and four vertical settings. You should
perform this procedure whenever the accuracy test shows a variance larger than ±0.1 degree.
Follow this procedure to calibrate the digital level:
1
Turn on the level by pushing the
ON/OFF
button. Place the level on a flat surface free of dust or debris.
You can use any horizontal surface within 10 degrees of level and any vertical surface within 10 degrees
of plumb to perform the calibration. You must use the same surface(s) throughout the entire calibration
process. If at any time the calibration process fails or will not continue, you can turn the level off and on
and then restart the calibration procedure from
2
Press and hold the
HOLD
and
ALT ZERO
buttons simultaneously until
SUP
is displayed; this usually
takes three seconds. Release the buttons once
SUP
is displayed.
SUP
will display for a few seconds
and then the
[ 0 ]
is displayed. The level is now ready to be calibrated.
Note
Each time you position the level during the calibration process, wait a minimum
of 10 seconds before pressing the
HOLD
button to advance to the next step.
Also ensure that the level is placed in the same position each time it is moved.
Содержание Scalar i6000
Страница 1: ...6 66881 01 A Scalar i6000 Installation Guide Installation Guide Installation Guide Installation Guide...
Страница 20: ...8 About This Guide and Your Product...
Страница 38: ...26 Installing a Stand Alone Control Module...
Страница 104: ...92 Installing a Multi Module Library...
Страница 156: ...144 Installing Cartridges...
Страница 164: ...152 Setting up Your Library for Access...
Страница 242: ...230 Configuring the Library...
Страница 304: ...292 Adding Optional Hardware eight 2 5 mm screws...
Страница 318: ...306 Adding Optional Hardware...
Страница 336: ...324 Installation Testing and Verification Figure 35 Example Test Log Output...
Страница 356: ...344 Testing and Calibrating the Digital Level...
Страница 362: ...350 LBX Board and Terminator...
Страница 380: ...368 Glossary...
Страница 384: ...372 Index...