![QuadTech Sentry Plus Series Скачать руководство пользователя страница 81](http://html2.mh-extra.com/html/quadtech/sentry-plus-series/sentry-plus-series_instruction-manual_3373286081.webp)
Service & Calibration
Page 81 of 85
4.3.5
IR Voltage Calibration (IRV)
Keep the setting on the Valhalla voltmeter at DC and 2kV range. IR calibration is applicable to
the Sentry 30 Plus instrument.
Press [F1] = UP to go to
Cal Step 5: ‘IRV 1kV Offset (50V)’
.
Press [STOP] to return instrument to steady state.
Press [START] to get offset value.
Press [F1] = INC or [F2] = DEC until the Sentry Plus display reads the same as the voltmeter.
Press [ENTER] to accept reading.
Sentry Plus reverts to “IRV Offset” display (incorporating new cal value).
Press [F1] = UP to go to
Cal Step 6: ‘IRV 1kV Full (1kV)’
. Change voltmeter range to 20kV.
Press [STOP] to return instrument to steady state.
Press [START] to get full value.
Press [F1] = INC or [F2] = DEC until the Sentry Plus display reads the same as the voltmeter.
Press [ENTER] to accept reading.
Sentry Plus reverts to “IRV Full” display (incorporating new cal value).
Содержание Sentry Plus Series
Страница 2: ...Page 2 of 85...
Страница 6: ...Page 6 of 85...
Страница 10: ...Page 10 of 85...
Страница 12: ...Page 12 of 85...
Страница 14: ...Page 14 of 85...
Страница 26: ......
Страница 66: ......
Страница 76: ......