Operation
Page 53 of 85
2.13
Connection to Device under Test
Figure 2-5 illustrates the connection of the Sentry Plus Series unit to a single DUT using the S02
1-meter HV cable set that comes standard with the instrument. The custom white banana
plug/red alligator clip is connected between the OUTPUT terminal on the Sentry Plus Series unit
and the high side of the device under test. The black banana plug/alligator clip is connected
between the RTN/LOW terminal on the Sentry Plus Series unit to the low side of the DUT.
DUT
S02
Cable
Set
+
-
F1
RTN/LOW
F2
OUTPUT
DANGER
PASS
FAIL
F1
F2
F3
F4
START
STOP
Sentry 30 Plus AC/DC/IR Hipot Tester
0
1
Q
uadTech
!
CAUTION
F4
F3
Max 5kVAC
6kVDC
CAL
UPDATE
1.250kV
STEP 1/5
STAND BY
OFST
0.500mA
3.0s
AC
LOW: OFF
ARC: OFF
RAMP: OFF
FALL: OFF
PROGRAM
PRESET
MENU
M1
RMT
LOCK
ERR
MORE..
Figure 2-5: S02 Cable Connection
Refer to paragraphs 3.2 through 3.7 for the description and illustration of the connection of
several QuadTech accessories to the Sentry Plus Series instrument.