POWER UP and STAND BY Displays
The Sentry Plus Series instruments have multiple menus or displays that may seem confusing at
first glance. In an attempt to clarify the numerous functions of the software, this instruction
manual will illustrate these displays in a (hopefully) logical format. The function keys (F1, F2,
F3 & F4) perform different tasks depending upon the menu currently shown on the display.
Figure 2-2 illustrates the instrument display when the instrument is initially turned on.
Figure 2-2: Power-Up Display
After 3 seconds, the instrument display reverts to the STAND BY display as illustrated in Figure
2-3. The box in the lower left hand corner denotes the instrument status.
Figure 2-3: STAND BY Display
To access the programming function of the Sentry Plus Series instrument in the STAND BY
menu, press the [F1] key (PROGRAM). Once in the PROGRAM display, select the test step
then the test mode (AC, DC, IR, GC or PA). Paragraphs 2.4 - 2.8 illustrate how to program the
specific parameters of each of the 5 tests.
NOTE
This manual is set up so if you follow the instructions in paragraphs 2.4-2.9 you will have programmed a 5-step test
(GC, AC, DC, IR and PA) and saved to memory location 1.
To enter programming mode.
To view/change preset (initial)
test parameters.
To view/change system parameters:
memory, system, option, calibration, key
lock, password, error, & about.
To view the programmed test setups and
access offset function.
Instrument Status
Remote
Status
On/Off
Front
Panel
Lock Out
Status
On/Off
Offset
Status
On/Off
Error
Status
On/Off
When box is highlighted (back lit),
the function is
ON.
PROGRAM
PRESET
MORE..
MENU
RMT
ERR
OFST
LOCK
STAND BY
1.250kV
3.0s
15.00mA
F1
F4
F3
F2
FALL: OFF
RAMP: OFF
ARC : OFF
LOW : OFF
AC
STEP 1/1
Function
of
F1 - F4 Keys
Test
Settings
Test Mode
AC, DC, IR,
GC or PA
Test Step 1/x
x = 1-10
Test Voltage (kV)
Test Time (sec)
Test Current (mA)
Memory
Location M
x
x
= 1-60
M1
STAND BY
EXIT
RMT
ERR
OFST
LOCK
WAIT 3s
F1
F4
F3
F2
Function of F1 - F4 Keys
Software
Memory
Line Frequency
Calibration
Model #
SENTRY PLUS 30 VERSION X.XX
Power Up Display: Instrument Initialization
DATA MEMORY CHECK
CALIBRATION DATA
AC LINE FREQUENCY
PASS
PASS
60 Hz
:
:
:
Содержание Sentry Plus Series
Страница 2: ...Page 2 of 85...
Страница 6: ...Page 6 of 85...
Страница 10: ...Page 10 of 85...
Страница 12: ...Page 12 of 85...
Страница 14: ...Page 14 of 85...
Страница 26: ......
Страница 66: ......
Страница 76: ......