Page 24 of 85
Introduction
1.4 Installation
1.4.1 Dimensions
The Sentry Plus series unit is supplied in a bench configuration, i.e., in a cabinet with resilient
feet for placement on a table. Flip feet are provided under the front feet so that the Sentry Plus
instrument can be tilted up for convenient operator viewing.
4.50"
112.50mm
14.25"
356.25mm
10.50"
262.5mm
F1
RTN/LOW
F2
OUTPUT
DANGER
PASS
FAIL
F1
F2
F3
F4
START
STOP
Sentry 30 Plus AC/DC/IR Hipot Tester
0
1
Q
uadTech
CAUTION
F4
F3
Max 5kVAC
6kVDC
CAL
UPDATE
Figure 1-4: Sentry Plus Series Instrument Dimensions
1.4.2
Instrument Positioning
The Sentry Plus unit contains a graphic display for direct readout of measured parameters. The
optimum angle for viewing is slightly down and about 10 degrees either side of center. For
bench operation the front flip feet should always be used to angle the instrument up. In bench or
rack mount applications the instrument should be positioned with consideration for ample air
flow around the rear panel fan ventilation hole. An open space of at least 3 inches (75mm) is
recommended behind the rear panel. Testing should be performed on a non-conductive surface.
An ESD mat is not a recommended test platform.
1.4.3
Power Requirements
The Sentry Plus instrument can be operated from a power source of 90 to 132V AC or 198 to
250V AC. Power connection is via the rear panel through a standard receptacle. Before
connecting the 3-wire power cord between the unit and AC power source, make sure the voltage
selection switches on the rear panel (Figure 1-5) are in accordance with the power source being
used. For a 90-132V source, use a 3.15A 250V fuse. For a 198-250V source, use a 1.6A 250V
fuse. Always use an outlet that has a properly connected protection ground.